OD Monitor control unit ODBox B
OD Monitor control unit ODBox B
It is a control unit of OD-MonitorB & L.
Main Application
●Cell density observation of bacterial in culturing●Cell density observation of pathogenic bacteria
●Cell observation of bacillus in culturing under anaerobic conditions which gas concentration is important
●Cell density observation of bacteria with slow multiplication rate
●A variety of seed culture
●Effective mass production of proteins by E.Coli etc.
Features
The measurement of even high concentration.
This unit uses "Hybrid OD method " which measures zero-point at start or low concentration range by transmitted light and high concentration range by reflected light from bacterial cells. It uses same principles with OD-MonitorA&S about the influence of fluid motion on measurement,.Set the Erlenmeyer flask size and shaking speed at the measurement.
Data can be saved in USB memory.
Automatic addition of reagent like IPTG.
If it is connected with an optional “Automatic pump unit”, IPTG etc. is added automatically at set OD value. (Automatic pump unit has 1ch each. Multi-connection is available.)
It can correct the gap of your spectrophotometer.
OD value may change by the measurement unit.The correction function makes the gap of your spectrophotometer within 15%.
Up to 4 sensors(option) can be connected.
Can measure under the same conditions by up to 4 sensors.Specifications
Model | ODBox-B |
Display of (OD) value | Display of converted values from transmitted light volume of infrared ray (950nm) to OD600 of E.coli bacteria |
Measurement Range | 0.1 - 9.9(accuracy guarantee range 0.3 - 7.0) |
Correction for biological species other than E. coli bacteria and gap from spectrophotometer | Enter correction factors (0.01 - 2.55、initial value is 1.00) |
Interval | 1/5/10/30/60/120/240min |
Maximum Measurement Time | Internal memory : 360 points (*1) USB memory : Up to USB memory |
Maximum Connectable ODSensor-L | 4sets Measurement under all same conditions |
Interface for external storage medium | Port for USB2.0 flash memory (*2) |
Other Functions | Alarm by buzzer when over target OD and activate pump (option) |
Dimensions | 153 x 220 x 127Hmm |
(*1)Point = Total measurement time / Interval. The 360th point figure is to updated when it exceed 360 points.
(*2)Some memory cannot be used with this unit even if it can be used with a personal computer.
(*2)Some memory cannot be used with this unit even if it can be used with a personal computer.
●The design and specifications are subject to change.