Unit water bath Thermominder SX-10N
Unit water bath Thermominder SX-10N
Higi-spec types of Thermominder.
Image of using "Stainless steel thermal insulation water bath" and "Heat-proof plastic hood" (option) |
Main Application
●Various incubation such as enzyme reaction●Inactivation of serum and thermal insulation of culture medium
●Hybridization
Features
SX-10N can memorize one temperature settings.
SX-10N can memorize one temperature setting, and this setting can use combination with timer.Temperature settings can change at a set time (1min - 99h59min).
Addressed to use at below room temperature
Temperature setting range is -20ºC - +105 ºC for SX-10N. Temperature control below room temperature enabled by combining with optional immersion cooler.Regulating strong and weak agitation is possible
Equipped with volume control which can control water jet arbitrarily. It can prevent ruffle when using small water bath or small volume of water.Variable heater capacity
Heater capacity can be changed between 0-100%.A lot of options other than attached plastic water bath
It is possible to facile immediate use with attached poly C water bath or optimization with options for container and purpose.Specifications
Model | SX-10N |
Temperature Range | RT+5ºC - +105 ºC(*) |
Temp. Control Accuracy | from ±0.05ºC |
Minimum temperature setting unit | 0.1ºC (setting and display) |
Temperature setting | By UP/ DOWN key |
Flow | Jet blast (volume adjust) |
Temperature Display | Digital display (switching between setting temperature and inside chamber temperature) |
Timer | Buzzer timer : 1min - 99h59min OFF timer : 1min - 99h59min Timer for change temperature setting : 1min - 99hr59min |
Memory | One temperature setting |
Heater | 1.0kW (SX-10N can time proportioning variable output: 0-100%) |
Weight | Approx. 3.4kg |
Water level range | Minimum water level: 80mm (from the bottom of PolyC water bath) |
Safety device / function | Circuit protector (10A), fuse (2A), float type prohibition of boiling while empty, protective cover for heater Misalignment high and low temperature error, lowing water, sensor error, auto tuning error, non-volatile storage error ※Notified by buzzer |
Occupying dimensions inside bath | 130 x 66 x 135Hmm |
Dimensions | 130 x 135 x 304Hmm |
Power Supply | AC100V 10.5A |
Standard Accessory | PolyC water bath ×1 , micro tube floater ×1 |
(*)Temperature setting range is -20ºC - +105 ºC for SX-10N.Temperature control below room temperature enabled by combining with optional immersion cooler.