TR-D-50
Application:Designed for precision liquid handling such as laboratory precision dilution,standard curve making and standard sample preparation, biological agents precise dosing, etc.
TR-D-50 Automatic Diluter is a high precision laboratory analyzer with Dilution, calibration, titration, filling function,it helps you to reduce the burden form using the pipette, volumetric flask or other accurate measuring vessel, no need to clean a lot of experimental equipment.
FEATURES
- The precise technology of constant volume supports a wide volume range from 0.4 mL to 3000 mL, and the minimum resolution reaches 0.01mL.
- The maximal dilution ratio reaches up to 7500, meeting various requirements of our users.
- The relative standard deviation of the precision is only 0.1% while the target volume is 100 mL.
- The relative error is less than 0.1%, and the accuracy is much higher than the Class A volumetric flask and manual dilution.
Comparison table of permissible error (According to JJG 196-2006
, Verification Regulation of Working Glass Container)
|
|||||||
Designated volume/mL
|
25
|
50
|
100
|
200
|
250
|
500
|
1000
|
Limit of error /mL ,
Class A Volumetric
Glassware
|
±0.03
|
±0.05
|
±0.10
|
±0.15
|
±0.15
|
±0.25
|
±0.40
|
The maximum relative
tolerance of Class A Volumetric Glassware
|
0.12%
|
0.10%
|
0.10%
|
0.075%
|
0.06%
|
0.05%
|
0.04%
|
The maximum relative
tolerance of D-50
|
0.08%
|
0.08%
|
0.06%
|
0.07%
|
0.05%
|
0.04%
|
0.035%
|
STANDARDS CONFIGURATION
TR-D-50 Instrument
|
1/PCS
|
Adapter
|
1/PCS
|
Accessories
|
1/Suit
|
Pump Pipe suit
|
2/Suit
|
User manual
|
1/PCS
|
100ml beaker
|
1/PCS
|
Temperature probe
|
1/PCS
|